DC Voltage
| Ranges |
Resolution |
Accuracy |
| 4V |
0.001V |
± (0.8% of reading + 3) |
| 40V |
0.01V |
| 400V |
0.1V |
| 600V |
1V |
± (1.0% of reading + 3) |
– Maximum input voltage: 600V DC RMS
-Input impedance: 10MΩ
AC Voltage
| Ranges |
Resolution |
Accuracy |
| 4V |
0.001V |
± (1.0% of reading + 10) |
| 40V |
0.01V |
| 400V |
0.1V |
| 600V |
1V |
± (1.2% of reading + 10) |
-Input impedance: 10MΩ.
-Over load Protection: 600V DC or 600V AC (RMS),
– Maximum input voltage: 600V AC (RMS)
-Response: True RMS
-Frequency Response: 40Hz~1kHz
Note: AC/DC voltage
At low voltage ranges, the meter may display fluctuating readings when probes are not connected to the circuit. This is normal and results from the meter’s high sensitivity. Connecting probes to the circuit will yield accurate measurements.
AC CURRENT(50Hz-1000Hz)
| RANGE |
RESOLUTION |
ACCURACY |
| 4A |
0.001A |
±(2.5% + 10 )
|
| 40A |
0.01A |
| 400A |
0.1A |
| 600A |
1A |
Maximum input current: 600A / Frequency range: 50–60Hz /
Response: True RMS
When measuring AC current, position the conductor to be tested at the center of the jaws. If not centered, positional error may increase by up to 1.5%.
RESISTANCE
| RANGE |
RESOLUTION |
ACCURACY |
| 400Ω |
0.1Ω |
±(1.2% of reading + 3)
|
| 4kΩ |
0.001kΩ |
| 40kΩ |
0.01kΩ |
| 400kΩ |
0.1kΩ |
| 4MΩ |
0.001MΩ |
| 40MΩ |
0.01MΩ |
±(2.0% of reading + 5) |
Overload protection: 250V DC/AC
FREQUENCY :
| RANGE |
RESOLUTION |
ACCURACY |
| 4Hz |
0.001Hz |
±(2.5% of reading + 10) |
| 40Hz |
0.01Hz |
| 400Hz |
0.1Hz |
| 4kHz |
0.001kHz |
| 40KHz |
0.01kHz |
| 400KHz |
0.1kHz |
| 4MHz |
0.001MHz |
Over load Protection: 250V DC or AC
Measurement Signal: Vpp 30mV
AC Signal: Note: As the measurement frequency increases, the Vpp input signal also increases.
CAPACITANCE
| RANGE |
RESOLUTION |
ACCURACY |
| 4nF |
0.001nF |
±(4.0% of reading + 5) |
| 40nF |
0.01nF |
±(4.0% of reading + 3) |
| 400nF |
0.1nF |
| 4µF |
0.001µF |
| 40µF |
0.01µF |
| 400µF |
0.1µF |
| 4mF |
0.001mF |
| 40mF |
0.01mF |
Overload protection: 250V DC or AC
DIODE & CONTINUITY
| Function |
Note |
| Diode |
Display approximate forward voltage drop of diodes |
| On/Off |
Built-in buzzer sounds when tested resistance is below 50 Ω |
| NCV |
Indicates signal strength based on distance from power source with varying beep frequencies |
| LIVE |
Displays “H” and sounds alarm when testing live wire; no response for neutral wire |
| POWER OFF |
Automatically powers off after 15 minutes
without signal input |